Cleaning, Sterilization, Inspection
and Maintenance of Reusable
Stainless Steel Surgical Instruments
Instructions for Renaissance Surgical, Inc.

Cleaning Recommendations:

Upon receipt inspect the instrument for defects. Carefully clean before sterilizing in preparation for its initial use. Always inspect the instrument for broken parts before each use. Clean and disinfect instrument immediately after use. Do not allow biofluids or other material to dry on the instrument: wipe clean and keep moist until the instrument reaches sterile processing. Rinse thoroughly in lukewarm water. Carefully scrub parts with soft brushes. Flush lumens and joints using a syringe filled with demineralized water. Use low-foaming, enzymatic cleaning solutions with a pH range of 6-8 and follow manufacturer’s instructions. Rinse with demineralized water. Lubricate in a water-soluble instrument bath. Store this instrument in a moisture-free location. Always cover and protect sharp edges and delicate tips when not in use. Sterilizing Recommendations: Upon receipt inspect this instrument for defects. Carefully clean before sterilizing in preparation for its initial use. Always inspect the instrument for broken parts before each use. Clean and disinfect instrument immediately after use. Do not allow biofluids or other material to dry on the instrument: wipe clean and keep moist until the instrument reaches sterile processing. Rinse thoroughly in lukewarm water. Carefully scrub parts with soft brushes. Flush lumens and joints using a syringe filled with demineralized water. Use low-foaming, enzymatic cleaning solutions with a pH range of 6-8 and follow manufacturer’s instructions. Rinse with demineralized water. Lubricate in a water-soluble instrument bath. Store this instrument in a moisture-free location. Always cover and protect sharp edges and delicate tips when not in use.

Sterilizing Recommendations:

METHOD

Steam (wrapped)

Steam (wrapped)

Steam (wrapped)

CYCLE

Prevacuum

Gravity

Gravity

MINIMUM TEMP

270°F (132°C)

270°F (132°C)

250°F (121°C)

EXPOSURE TIME

4 minutes

10 minutes

45 minutes

Due to the chemical sterilization and hot air, or steam sterilization and hot air, steam sterilization in excess of 280°F (138°C) should be avoided.
Allow the instrument to dry for a minimum of 20 minutes prior to using in surgery.

Document #54791345-22336